Neutron dosimetry using aqueous solutions of lithium acetate

1

TL;DRAbstract

A thermal neutron dosimetry system using the 6Li(n,α)3H reaction and liquid scintillation counting of tritium was developed. Lithium acetate was chosen to supply the 6Li in the aqueous dosimetry solutions. Neutron irradiations were completed using The Ohio State University Research Reactor. After two sets of samples were irradiated, variables in the system such as the mass of lithium acetate in the solutions and the counting window of the liquid scintillation counter used to analyze the sample were chosen. The system was evaluated by completing two sets of 23 minute irradiations with the reactor at 500 kW, 50 kW, 5 kW, and one irradiation at 500 W. The samples irradiated at 500 W were below the threshold of the system, and could not be used. Prompt analysis was essential due to loss of detectable emissions in the dosimetry solutions over time. The thermal neutron fluences calculated with the data from the samples were compared to the fluences determined from gold wire irradiations. The

Chat with Paper

AI Agents for this Paper

A thermal neutron dosimetry system using the 6Li(n,α)3H reaction and liquid scintillation counting of tritium was developed. Lithium acetate was chosen to supply the 6Li in the aqueous dosimetry solutions. Neutron irradiations were completed using The Ohio State University Research Reactor. After two sets of samples were irradiated, variables in the system such as the mass of lithium acetate in the solutions and the counting window of the liquid scintillation counter used to analyze the sample were chosen. The system was evaluated by completing two sets of 23 minute irradiations with the reactor at 500 kW, 50 kW, 5 kW, and one irradiation at 500 W. The samples irradiated at 500 W were below the threshold of the system, and could not be used. Prompt analysis was essential due to loss of detectable emissions in the dosimetry solutions over time. The thermal neutron fluences calculated with the data from the samples were compared to the fluences determined from gold wire irradiations. The

Keywords

DosimetryIrradiationRadiochemistryResearch reactorFluenceNeutron temperatureLiquid scintillation countingLithium (medication)

Chat

Click to start Chat