CitedEvidence
User Settings

WE-D-BRA-01: FEATURED PRESENTATION and BEST IN PHYSICS (THERAPY): Predicting Potentially Problematic VMAT Treatment Plans Before Patient Specific QA Measurements

0

TL;DRAbstract

Purpose: Several promising IMRT QA tools have been developed in recent years to combat problems found in the lack of sensitivity in planar dose measurements analyzed using consensus gamma analysis criteria. The increased complexity and added information with such devices adds not only increased time, but new challenges in determining endpoints for pass/fail criteria. Using a large cohort of previously measured planar IMRT QA data, it may be possible to correlate potentially problematic plans with calculated plan metrics that can be done a priori, such that these tools can be used only in clinically relevant situations. Methods: 90 previously measured, clinically delivered VMAT plans were exported in DICOM RT format. Using a Matlab program, plan metrics were computed based on a previously developed set of equations (Du et al. 2014). These metrics included MU-weighted beam irregularity, which quantifies an MLC shape's deviation from that of a circle. Machine delivery parameters such as M

Chat with Paper

AI Agents for this Paper

Purpose: Several promising IMRT QA tools have been developed in recent years to combat problems found in the lack of sensitivity in planar dose measurements analyzed using consensus gamma analysis criteria. The increased complexity and added information with such devices adds not only increased time, but new challenges in determining endpoints for pass/fail criteria. Using a large cohort of previously measured planar IMRT QA data, it may be possible to correlate potentially problematic plans with calculated plan metrics that can be done a priori, such that these tools can be used only in clinically relevant situations. Methods: 90 previously measured, clinically delivered VMAT plans were exported in DICOM RT format. Using a Matlab program, plan metrics were computed based on a previously developed set of equations (Du et al. 2014). These metrics included MU-weighted beam irregularity, which quantifies an MLC shape's deviation from that of a circle. Machine delivery parameters such as M

Keywords

DICOMOutlierNuclear medicineMedical physicsComputer sciencePlan (archaeology)Radiation treatment planningReceiver operating characteristic

Chat

Click to start Chat