SU‐E‐T‐95: An Alternative Option for Reducing Lung Dose for Electron Scar Boost Irradiation in Post‐Mastectomy Breast Cancer Patients with a Thin Chest Wall

Y Lee,P. Kumar,Melissa P. Mitchell-2015-06-01-Medical Physics
0

TL;DRAbstract

Purpose: Breast cancer patients who undergo a mastectomy often require post‐mastectomy radiation therapy (PMRT) due to high risk disease characteristics. PMRT usually accompanies scar boost irradiation (10–16Gy in 5–8 fractions) using en face electrons, which often results in increased dose to the underlying lungs, thereby potentially increasing the risk of radiation pneumonitis. Hence, this study evaluated water‐equivalent phantoms as energy degraders and as an alternative to a bolus to reduce radiation dose to the underlying lungs for electron scar boost irradiation. Methods: Percent depth dose (PDD) profiles of 6 MeV (the lowest electron energy available in most clinics) were obtained without and with commercial solid water phantoms (1 to 5mm by 1mm increments) placed on top of electron cones. Phantom attenuation was measured by taking a ratio of outputs with to without the phantoms in 10×10cm2 cone size for monitor unit (MU) calculation. In addition, scatter dose to contralateral b

Chat with Paper

AI Agents for this Paper

Purpose: Breast cancer patients who undergo a mastectomy often require post‐mastectomy radiation therapy (PMRT) due to high risk disease characteristics. PMRT usually accompanies scar boost irradiation (10–16Gy in 5–8 fractions) using en face electrons, which often results in increased dose to the underlying lungs, thereby potentially increasing the risk of radiation pneumonitis. Hence, this study evaluated water‐equivalent phantoms as energy degraders and as an alternative to a bolus to reduce radiation dose to the underlying lungs for electron scar boost irradiation. Methods: Percent depth dose (PDD) profiles of 6 MeV (the lowest electron energy available in most clinics) were obtained without and with commercial solid water phantoms (1 to 5mm by 1mm increments) placed on top of electron cones. Phantom attenuation was measured by taking a ratio of outputs with to without the phantoms in 10×10cm2 cone size for monitor unit (MU) calculation. In addition, scatter dose to contralateral b

Keywords

Imaging phantomBolus (digestion)MedicineNuclear medicineRadiation therapyBreast cancerMastectomyDosimetry

Chat

Click to start Chat